Krautkramer Phasor Series

SKU:
$2,997.00
$2,997.00
Unavailable
per item
The Phasor series ultrasonic flaw detectors comes available in four different models with upgraded features and benefits. Phasor CV is a code-compliant conventional ultrasonic flaw detector that can be UPGRADED to Phasor 16/16 Weld or to the 16/64 Phasor XS. Upgrading to phased array is a quick and easy software upgrade to Phasor 16/16 Weld or Phasor XS. No changes to the base unit hardware are required and utilizing the same simple menu-driven interface as the Phasor CV model, training is minimal. Combined with phased array transducers, both platforms can solve your most demanding inspection applications in less time at an affordable price.
Phasor 16/16 Weld and Phasor XS can switch to conventional UT quickly so that inspectors can use any standard transducer to evaluate detected flaws (locating and sizing) to any specified standard or test instruction.
The Phasor CV/DM combines a phased array imaging device and a conventional flaw detector into one instrument, with a phased array probe, which allows significant increases confidence in corrosion inspection data

Spectification:
Internal Memory    Set-up files
Removable Memory    On 512 MB SD Card for report and set-up files
Documentation Format    JPEG ~80 KB/image
Weight    3.8 kg (8.2 lbs) with battery
Dimensions    282 mm W x 171 mm H x 159 mm D (11.1 in. W x 6.8 in. H x 6.3 in. D)
Battery    Custom Li Ion battery pack - 356P configuration
Battery Life    6 hrs minimum
Battery Charging    External charger
External Power Supply    Universal input 85 to 260 V AC / 50 to 60 Hz
Probe Connectors    Conventional - 00 lemo/BNC adapters provided - Phased Array - Custom ZIF
VGA Output    Yes
Dialog Languages    Chinese, English, French, German, Italian, Japanese, and Spanish
Display Size    165 mm (6.5 in.) diagonal
Display Resolution    VGA color TFT 640H x 480V pixel
Conventional / Phased Array Channel

Conventional    Phased Array
Pulser    Spike    Bi-Polar Square Wave
Pulse Repetition Frequency    15 to 2000 Hz    15 to 7680 Hz
Pulser Voltage    300 V max    ± 25 V to ± 75 V (1 V steps)
Pulser Energy    Low or High (selectable)   
Pulser Rise Time    < 15 nsec    < 15 nsec
Damping    50 or 1000 Ohms (selectable)   
Mode of operation    single, dual    single
Receiver Input Capacitance    < 50 pF   
Receiver Input Resistence    1000 Ohms in dual mode    220 Ohms
Maximum Input Voltage    40 V peak-to-peak    200 mV peak-to-peak
Bandwidth/Amplifier Bandpass    0.3 to 15 MHz @ -3dB    0.6 to 6.5 MHz @ -3dB
Filter Selection    1.0, 2.0. 2.25, 4.0, 5.0, 10 and 13 MHz + BB    2.0, 3.0, 4.0, 5.0 MHz, Low Pass 4 MHz, High Pass 5 MHz and Broadband (0.6 to 6.5 MHz)
Rectification    Pos HW, Neg HW, FW, and RF    Pos HW, Neg HW, FW and RF
Analog Gain    0 to 110 dB    0 to 40 dB
Digital Gain 1        5.8 to 39.9 dB (depending on aperturesize)
Focal Laws        User selectable - 128 max
Physical Probe        1 to 64
Virtual Probe        1 to 16
Number of Cycles        1 to 128
Pulser Width @ 1/2 Cycle        20 to 500 nsec
Pulser Delay

0 to 10.24 µ-sec
Receiver Delay   
0 to 10.24 µ-sec
Acoustic Velocity    1000 to 16000 m/s
0.0393 to 0.5905 in./µ-sec
1000 to 16000 m/s
0.0393 to 0.5905 in./µ-sec
Minimum Range (steel long)
(steel shear)    0 -14 mm (0.55 inch)
0 - 7.5 mm (0.3 inch)
0 - 7.6 mm (0.3 inch)
0 - 4.2 mm (0.17 inch)
Maximum Range (steel long)
(steel shear)    0 - 14060 mm (553 inch)
0 - 7626 mm (300 inch)
0 - 1073 mm (42 inch)
0 - 1073 mm (42 inch)
Display Delay    2.5 m (98.5 inch)
1 m (39.4 inch)
Auto Timebase Calibration    Yes

Reject    0 to 80%
0 to 80%
TCG    15 points @ 6 dB/µ-sec
15 points @ 6 dB/µ-sec
Gates    A and B
A, B and IF
Gate Threshold    5 to 95%
5 to 95%
Gate Start    0 mm - full range
0 mm - full range
Gate Width    1 mm - full range
1 mm - full range
Gate Logic    Off, Positive, and Negative
(Off, Coincidence, and Anticoincidence)
Off, Positive, and Negative
(Off, Coincidence, and Anticoincidence)
TOF Modes    Flank/Peak
Flank/Peak
Scan Type   
Linear and Sector
Available Views    A-Scan
A-Scan, B-Scan and Sector
Displayed Readings    Amplitude, Sound Path, and Trig
Beam, Amplitude, Sound Path,
Trig for displayed and for all beams
Measurement Resolution    5 nsec
5 nsec
Displayed Units of Measurements    mm or inch (selectable)
mm or inch (selectable)
Environmental Tests (Per Mil-Std-810F)

Cold Storage    -20°C for 72 hrs, 502.4
Procedure I
Cold Operation    0°C for 16 hrs, 502.4
Procedure II
Heat Storage    +70°C for 48 hrs, 501.4
Procedure I
Heat Operation    +50°C for 16 hrs, 501.4
Procedure II
Damp Heat / Humidity
(storage)    10 Cycles:
10 hrs at +65°C down to
+30°C,
10 hrs at +30°C up to +65°C,
transitions within 2 hrs, 507.4
Temperature Shock    3 Cycles:
4 hrs at –20°C up to +70°C,
4 hrs
at +70°C, transitions within 5
mins.
503.4 Procedure II
Vibration    514.5-5 Procedure I, Annex C,
Figure 6, general exposure:
1 hr each axis
Shock    6 cycles each axis, 15g, 11ms
half sine, 516.5 Procedure I
Loose Cargo    514.5 Procedure II
Transit Drop
(packaged for shipment)    516.5 Procedure IV, 26 drops
IP54 / IEC529 … dust proof / dripping water proof as per
IEC 529 specifications for IP54 classification
Phasor CV/DM Option

Pulser    Bi-Polar Square Wave
Pulser Repetition Freq.    15 to 7680 Hz
Pulser Voltage    50 V to 150 Volts 2V Steps
Pulser Rise Time    < 15 nsec
Mode of Operation    Custom T/R operating files only
Receiver Input Resistance    220 Ohms
Max Input Voltage    200 mV peak-to-peak
Bandwidth/Amplifier
Bandpass    0.6 to 6.5 MHz @ -3 dB
Filter Selection    2.0, 3.0, 4.0, 5.0 MHz, Low Pass 4 MHz, High Pass
5 MHz and Broadband (0.6 to 6.5 MHz)
Rectification    Pos HW, Neg HW, FW and RF
Analog Gain    0 to 40 dB
Digital Gain    0 to 53.9 DB depending on APERTURE selection
Virtual Probe    1 to 16 elements
Pulser Width @ 1/2 Cycle    20 to 500 nsec
Pulser Delay    0 to 10.24 µsec
Receiver Delay    0 to 10.24 µsec
Acoustic Velocity    1000 to 16000 m/sec, 0.0393 to 0.5905”/µsec
Steel Min Range    Long: 0-7.6 mm (0.3”), Shear: 0-4.2 mm (0.17”)
Steel Max range    Long: 0-1073 mm (42”), Shear: 0-1073 mm (42”)
Display Delay    1m (39.4”)
DM T/R Option

TCG    15 points @ 6 dB/µsec
Gates    A, B and IF
Gate Threshold    5 to 95%
Gate Start    0 mm - full range
Gate Width    1 mm - full range
Gate Logic    Off, Positive and Negative (Off, Coincidence and
Anticoincidence)
TOF Modes    Flank/Peak
Available Views    ASCAN, BSCAN and TOP VIEW
Displayed Readings    Beam, Amplitude, Sound Path, Trig for displayed
and for all beams
Measurement Resolution    5 nsec
Displayed Units    mm or inch (selectable)
DM Option Specifications SW 2.27 Corrosion

Measuring modes    Zero to GATE crossing, Flank or Peak and Gate to Gate
Calibration    2 point Manual Probe zeroing and material CAL,
probe wear compensation
Measuring    V-path correction
Thin Reading Capture
modes    Per Beam minimum, Per Scan minimum, Accumulated
minimum and TOPVIEW auto cursor location to minimum
Probe compatability    Compatible Probes (as of October 09)
Global PN 1314599 5Mhz 32element 1.5mm pitch
Global PN 1314699 5Mhz 32 element .75mm pitch
Temperature range –29°C (-20°F) to 57°C (135°F)
* See instrument and probe documentation for
important notes.